Question | Answer |
---|---|
AEC will produce consistant density regardless of | Patient Size or Pathological Condition |
Benefits of AEC | decrease pt dose, increase departmental efficiency, reduced film cost, reduced processing chemical cost, & prolonged tube life |
Two types of AEC units | ionization chamber & phototimer type. |
The ionization chambers in an AED are located ___ the cassette. | In front of |
All of the following are advantages of AEC, except | Eliminates need for proper positioning |
Minimum response time refers to the time that must be set by the radiographer when making a radiographic exposure | false |
The purpose of the backup timer in an AED is to | Protect the pt and Protect the x-ray tube |
With AEC, each incremental density setting corresponds to a ___ change in density | 25% |
With which of the AEC units listed below does backscatter have the greatest impact on image quality? | Phototimer type |
The two most important factors affecting film density when using AEC are | positioning and detector selection |
All of the following can cause overexposure when using AEC except | SID |
It is acceptable tp use the center detector when performing abdominal radigraphy | False |
The outer detectors can be used with large field sizes | True |
OID affects density when uing AEC | False |
Increasing kV and mA can shorten the minimum response time | False |
Beam intensity affects exposure time when using AEC | True |
Activation of the backup timer is most often caused by | Radiographic Error |
Backup timers should never allow the pt to be exposed to more than ___ mAs | 600 |
If the xray tube is not properly aligned to the Bucky grid, the exposure time will be | Excessive long |
The exposure time can be intentionally increased when using AEC by | Decreasing mA |
When using AEC, if a fast-speed sreen cassette were used instead of a medium-speed screen cassette, the exposure time would be | unaffected |
All of the following can affect AEC exposure time, except | Film speed |
Insufficient collimation __ exposure time when using AEC | Reduces |
The ionization chambers are situated between grid of the bucky device and the cassette. | (Usually 3 chambers) |
Two most important factors that affect image quality to consider are pt positioning and detector selection. |