ESE readout for cassette spot films in flouro: |
According to federal law ESE for flouro shall not exceed: |
10R per minute (100mGy per min) |
How much can the tolerance for reproducibility be off? |
+/- 5% (same as the variance in voltage supply) |
According to federal law ESE for angiography shall not exceed: |
20R per minute (200mGy per min) |
Allowable variance in tube arc for tomo (greater than 30 degrees): |
a sequential increase in mAs should produce the same sequential increase in exposure |
What is the 9 penny test used for? |
to determine if the alignment between the collimator and bucky is within range. |
Allowable variance in tube arc for tomo (less than 30 degrees): |
What filtration is needed for tubes operating at 50 kVp and below? |
How do we know if the rectifier is at fault for an exposure timer error? |
there should be 30 dots, half the number of dots always means it’s the rectifier |