img crit ch 33 Matching
ESE readout for cassette spot films in flouro:
200mR
According to federal law ESE for flouro shall not exceed:
10R per minute (100mGy per min)
How much can the tolerance for reproducibility be off?
+/- 5% (same as the variance in voltage supply)
According to federal law ESE for angiography shall not exceed:
20R per minute (200mGy per min)
Allowable variance in tube arc for tomo (greater than 30 degrees):
+/- 5 degrees
Linearity
a sequential increase in mAs should produce the same sequential increase in exposure
What is the 9 penny test used for?
to determine if the alignment between the collimator and bucky is within range.
Allowable variance in tube arc for tomo (less than 30 degrees):
+/- 2 degrees
What filtration is needed for tubes operating at 50 kVp and below?
.5 mmAleq
How do we know if the rectifier is at fault for an exposure timer error?
there should be 30 dots, half the number of dots always means it’s the rectifier