Equipment Operation and Quality Control

Quiz yourself by thinking what should be in each of the black spaces below before clicking on it to display the answer.
        Help!  

Question
Answer
True or False? The effective of projected focal spot is always smaller than the actual focal spot.   True pg. 296  
🗑
What functions to protect the patient from overexposure and the x-ray tube from overload?   back up timer pg. 298  
🗑
What describes the chemical ionization effects of an electric current?   electrolysis pg. 298  
🗑
What type of relationship do kV and HVL have?   direct relationship pg. 300  
🗑
True or False? The step down transformer is located in the filament circuit and serves to regulate the voltage and current provided to heat the x-ray tube filament.   True pg. 300  
🗑
Decreasing SID increases? or decreases? the intensity of the x-ray photons at the image intensifiers input phosphor?   increases pg. 300  
🗑
What are the three types of beam restrictors?   aperature diaphragms, cones and cylinders, collimators pg. 302  
🗑
Electrons are focused carefully to maintain image resolution by what device?   electrostatic focusing lenses pg. 303  
🗑
A charged coupling device converts visible light to an electrical charge that is sent to what for processing?   analog to digital converter pg. 303  
🗑
PBL must be accurate to what percent of the SID?   2% pg. 305  
🗑
Reproducibility specifies that radiation output must be consistent to within what percent?   +/- 5% pg. 310  
🗑
What are the two main parts of an induction motor?   stator and rotor pg. 311  
🗑
What functions to direct the liberated filament electrons to the focal spot?   Molybdenum focusing cup pg. 312  
🗑
Which type of mobile x-ray unit is capable of storing a large milliampere seconds capacity for extended periods of time?   battery powered pg. 312  
🗑
True or False With 3 phase, 12 pulse equipment, the voltage never drops to 0.   True pg. 313  
🗑
What device is used to protect the circuit elements from overload by opening the circuit in the event of a power surge?   fuse pg. 314  
🗑
With single phase equipment, the number of heat units produced during a single exposure is determined using what equation?   mA x time x kV pg. 315  
🗑
Three phase, twelve pulse equipment has what percent voltage ripple?   3.5% pg. 315  
🗑
The glass envelope and oil coolant provide what type of filtration?   inherent pg. 316  
🗑
What is used to test timer accuracy on three phase equipment?   oscilloscope or synchronous spinning top pg. 317  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: grbuechler
Popular Radiology sets