Ch. 18-23, 25, 26
Quiz yourself by thinking what should be in
each of the black spaces below before clicking
on it to display the answer.
Help!
|
|
||||
---|---|---|---|---|---|
show | lumen
🗑
|
||||
A digital display monitor is best viewed | show 🗑
|
||||
A liquid crystal is a material in a state between a liquid and a | show 🗑
|
||||
Almost all digital images in medical imaging are viewed and interpreted on a | show 🗑
|
||||
Medical flat panel display devices are _______ liquid crystal displays. | show 🗑
|
||||
Medical flat panel display devices are identified by the number of | show 🗑
|
||||
Compared to CRT's, AMLCD's have better _____ and less intrinsic _______ | show 🗑
|
||||
The ambient light at a digital image workstation should be near | show 🗑
|
||||
show | automatically; manually
🗑
|
||||
Flatfielding is accomplished through automatic calibration images called ____ and ____ images. | show 🗑
|
||||
Reregistering an image to correct for patient motion is done with | show 🗑
|
||||
show | preprocessing
🗑
|
||||
The national standard for image transmission in teleradiology is the | show 🗑
|
||||
Image field size is determined by | show 🗑
|
||||
A screen film radiographic artifact is any irregularity on an image that is not caused by the proper imaging or tissue by primary | show 🗑
|
||||
show | exposure artifacts
🗑
|
||||
show | processor artifact
🗑
|
||||
show | exposure artifacts
🗑
|
||||
show | guide shoe marks
🗑
|
||||
Thin lines on the leading or trailing edge of a film are | show 🗑
|
||||
Only patients can cause image | show 🗑
|
||||
show | bending the film
🗑
|
||||
A quality assurance program monitors | show 🗑
|
||||
show | equipment
🗑
|
||||
show | outcome analysis
🗑
|
||||
show | assurance
🗑
|
||||
What does JCAHO mean | show 🗑
|
||||
show | ensure quality patient care, receive payment, JCAHO approval
🗑
|
||||
Most quality control testing on radiographic equipment is done | show 🗑
|
||||
show | half value layer measurements
🗑
|
||||
The collimator must be accurately aligned to within + or - | show 🗑
|
||||
The measure of kVp should be within + or - ___ of the kVp indicated on the console. | show 🗑
|
||||
show | collimation
🗑
|
||||
show | erasure
🗑
|
||||
_____ is done during preprocessing to correct for _____ over the image | show 🗑
|
||||
___ is done during the imaging, flatfielding is done during the preprocessing, and lossless compression is done during post processing | show 🗑
|
||||
CR cassettes can sit for a few _____ between exposures. | show 🗑
|
||||
Lossless compression reduces data files by a ratio of | show 🗑
|
||||
Lossy compression reduces data by a ratio of | show 🗑
|
||||
show | uncompressed images
🗑
|
||||
show | histogram
🗑
|
||||
Medical images are transferred using the | show 🗑
|
||||
Soft tissue radiography is unique because of the | show 🗑
|
||||
show | enhanced
🗑
|
||||
Breast tissue tends to be increasingly ___ in older women | show 🗑
|
||||
About ___ of breast cancer occurs in the ductal tissue | show 🗑
|
||||
The American Cancer Society recommends all women ___ years of age and above have an annual screening mammogram | show 🗑
|
||||
The ACS recommends biannual screening for all women age | show 🗑
|
||||
A dedicated mammography unit should have an automatic | show 🗑
|
||||
Every dedicated mammography unit is equipped with a | show 🗑
|
||||
show | patient does, motion blur, and superimposition
🗑
|
||||
show | reduce focal spot blur
🗑
|
||||
Cassettes used in mammography are designed with tighter | show 🗑
|
||||
show | 5 mA
🗑
|
||||
The image intensifier improved fluoroscopy by increasing image | show 🗑
|
||||
Visual acuity in the eye is greater at the fovea centralis where | show 🗑
|
||||
show | contrast perception
🗑
|
||||
The rods in the retina are stimulated by ___ light; the cones are stimulated by _____ light. | show 🗑
|
||||
show | input phospher
🗑
|
||||
show | light; cesium iodide
🗑
|
||||
The number of light photons emitted within the image intensifier is directly proportional to the amount of xray photons | show 🗑
|
||||
Light produced at the output phosphor has been increased | show 🗑
|
||||
show | flux gain
🗑
|
||||
show | brightness gain
🗑
|
||||
show | input phosphor
🗑
|
||||
show | 80 to 90 kVp
🗑
|
||||
Viewing the fluoroscopy image in magnification mode increase | show 🗑
|
||||
show | kVp and mAs
🗑
|
||||
Digital fluoroscopy uses at least | show 🗑
|
||||
The time it takes to turn on the digital fluoroscopy xray tube and reach the selected mA and kVp is the | show 🗑
|
||||
The charged coupled devise used in digital fluoroscopy provides | show 🗑
|
||||
Digital fluoroscopy dose rate is lower than that for continuous analog fluoroscopy because the images are | show 🗑
|
Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
To hide a column, click on the column name.
To hide the entire table, click on the "Hide All" button.
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.
To hide a column, click on the column name.
To hide the entire table, click on the "Hide All" button.
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.
Normal Size Small Size show me how
Normal Size Small Size show me how
Created by:
10153696170312934
Popular Radiology sets