image critique - QA and QC

        Help!  

Question
Answer
What filtration is needed for tubes operating at 70 kVp and above?   2.5 mmAleq  
🗑
What filtration is needed for tubes operating between 50 and 70 kVp?   1.5 mmAleq  
🗑
What filtration is needed for tubes operating at 50 kVp and below?   .5 mmAleq  
🗑
How much can the alignment of the collimator and bucky tray be off?   +/- 2% of the SID  
🗑
How much can the tolerance for the CR be off?   +/- 1%  
🗑
How much can the tolerance for the angulation be off?   +/- 1%  
🗑
How much can the tolerance for the exposure timer be off?   +/- 5% if over 10ms or +/-20% if under or at 10ms  
🗑
How do we check the phototimer accuracy?   expose 2 AL or acrylic pieces of different thickness. They should have the same OD  
🗑
How do we know if the rectifier is at fault for an exposure timer error?   there should be 30 dots, half the number of dots always means it’s the rectifier  
🗑
What is the 9 penny test used for?   to determine if the alignment between the collimator and bucky is within range.  
🗑
Reciprocity   combinations of mA and s that yield the same mAs  
🗑
Linearity   a sequential increase in mAs should produce the same sequential increase in exposure  
🗑
Reproducibility   exposure she be the same if the same settings are used  
🗑
How much can the tolerance for reproducibility be off?   +/- 5% (same as the variance in voltage supply)  
🗑
How much can the tolerance for film illuminators (viewboxes) be off?   +/- 10%  
🗑
According to federal law ESE for flouro shall not exceed:   10R per minute (100mGy per min)  
🗑
According to federal law ESE for angiography shall not exceed:   20R per minute (200mGy per min)  
🗑
ESE readout for cassette spot films in flouro:   200mR  
🗑
ESE readout for photo-flourospot images in flouro:   100mR  
🗑
Allowable variance in tube arc for tomo (greater than 30 degrees):   +/- 5 degrees  
🗑
Allowable variance in tube arc for tomo (less than 30 degrees):   +/- 2 degrees  
🗑
Repeat rates should fall between:   3-10% averaging at 8%  
🗑


   

Review the information in the table. When you are ready to quiz yourself you can hide individual columns or the entire table. Then you can click on the empty cells to reveal the answer. Try to recall what will be displayed before clicking the empty cell.
 
To hide a column, click on the column name.
 
To hide the entire table, click on the "Hide All" button.
 
You may also shuffle the rows of the table by clicking on the "Shuffle" button.
 
Or sort by any of the columns using the down arrow next to any column heading.
If you know all the data on any row, you can temporarily remove it by tapping the trash can to the right of the row.

 
Embed Code - If you would like this activity on your web page, copy the script below and paste it into your web page.

  Normal Size     Small Size show me how
Created by: annaluz87
Popular Radiology sets