| Question | Answer |
| What filtration is needed for tubes operating at 70 kVp and above? | 2.5 mmAleq |
| What filtration is needed for tubes operating between 50 and 70 kVp? | 1.5 mmAleq |
| What filtration is needed for tubes operating at 50 kVp and below? | .5 mmAleq |
| How much can the alignment of the collimator and bucky tray be off? | +/- 2% of the SID |
| How much can the tolerance for the CR be off? | +/- 1% |
| How much can the tolerance for the angulation be off? | +/- 1% |
| How much can the tolerance for the exposure timer be off? | +/- 5% if over 10ms or +/-20% if under or at 10ms |
| How do we check the phototimer accuracy? | expose 2 AL or acrylic pieces of different thickness. They should have the same OD |
| How do we know if the rectifier is at fault for an exposure timer error? | there should be 30 dots, half the number of dots always means it’s the rectifier |
| What is the 9 penny test used for? | to determine if the alignment between the collimator and bucky is within range. |
| Reciprocity | combinations of mA and s that yield the same mAs |
| Linearity | a sequential increase in mAs should produce the same sequential increase in exposure |
| Reproducibility | exposure she be the same if the same settings are used |
| How much can the tolerance for reproducibility be off? | +/- 5% (same as the variance in voltage supply) |
| How much can the tolerance for film illuminators (viewboxes) be off? | +/- 10% |
| According to federal law ESE for flouro shall not exceed: | 10R per minute (100mGy per min) |
| According to federal law ESE for angiography shall not exceed: | 20R per minute (200mGy per min) |
| ESE readout for cassette spot films in flouro: | 200mR |
| ESE readout for photo-flourospot images in flouro: | 100mR |
| Allowable variance in tube arc for tomo (greater than 30 degrees): | +/- 5 degrees |
| Allowable variance in tube arc for tomo (less than 30 degrees): | +/- 2 degrees |
| Repeat rates should fall between: | 3-10% averaging at 8% |