click below
click below
Normal Size Small Size show me how
THMG Drager XAM7000
Meter questions
Question | Answer |
---|---|
What is the calibration gas for the PID in this meter? | Isobutylene |
How long is the typical warm up time for this meter? | 3-4.5 minutes |
What is the maximum feet of tubing the pump is rated to pull a sample through? | 150 feet of tubing |
How long will the typical battery last in operation? | 14 hours |
For PID detection, what are the ranges? | 0-2000 ppm |
For PID detection, what is the low alarm? | 100 ppm |
For PID detection, what is the high alarm? | 200 ppm |
What is the power of the PID lamp? | 10.6 eV |
For EXplosive detection, what is the low alarm? | 10% |
For EXplosive detection, what is the high alarm? | 20% |
What is the range for the oxygen sensor? | 0-25% |
What is the low and high alarm levels for oxygen? | 19.5 - 23.5% |
What is the range for the CO sensor? | 0-2000 ppm |
What are the low and high ACTION LEVELS for CO? | 10 for low and 100 for the high |
What are the low and high ALARM LEVELS for the CO? | 30 ppm and 60 ppm |
What is the range for the H2S sensor? | 0-100 ppm |
What is the low alarm for the H2S sensor? | 10 ppm |
What is the high alarm for the H2S sensor? | 20 ppm |
How long is the startup sequence for the XAM7000? | 3.5 to 4.5 minutes |
Is the fresh air setup for the XAM7000? | No, unless the meter notes a sensor drift |
What does the up arrow do when the meter is off? | It will display info including the battery status |
What is the hose made of? | Polyurethane |
How do you use the meter in Diffusion Mode | Remove back plate where the magnets are held, thats what engages the pump mode |
How long does it take Air to move through 3 ft of hose? | 1 second |
What does an " ! " mean on the display screen? | Indicates a message or notice |
What does a " X " mean on the display screen? | Indicates there is an error or fault with a meter funciton |
What does 4 "up arrows" in a row mean on the display screen? | The Gas is above the Max Range for the sensor |
What does "- - - - " mean on the display screen? | Flow alarm while in pump mode /Pumps not drawing enough air |
On the Red Strap, what is the high capacity O2 range? | 0 - 100% |
What can cause faulty readings for O2 sensors? | High Humidity and High / Low Altitudes |
What is the range for the CO sensor on the Red Strap? | 0 - 10,000 PPM |
What is the range for H2S sensor on the Red Strap? | 0 - 1,000 PPM |
What is the main benefit for using IR (infrared) Sensors? | IR sensors work in O2 deficient atmospheres |
What is a downfall for using IR Sensors | IR sensors cannot see gases that are lighter than air, such as Acetylene and Hydrogen |
What Gas is the LEL Sensor calibrated to? | Methane |
What is the High CO2 Sensor's range and alarm levels on the Red Strap? | (0 - 100%) Low 1%; High 3% |