click below
click below
Normal Size Small Size show me how
RADT 456 Equip and Q
ARRT Registry Review
Question | Answer |
---|---|
What is the purpose of the filament current? | Supplying low-voltage current to the filament of the x-ray tube. (Saia, Radiology Prep: p. 455) |
What causes pitting or many small surface melts of the anode focal track? | Repeated and frequent overloading (Saia, Lange Q&A: p. 280) |
Congruence of the x-ray beam with the light field is tested by? | Radiopaque objects (Saia, Lange Q&A: p. 281) |
Which of the following is/are components of the secondary side of the x-ray circuit: 1. Rectification system 2. Autotransformer 3. kV meter | (1) Rectification system (Saia, Lange Q&A: p. 282) |
What device is used to convert electrical energy to mechanical energy? | Motor (Saia, Lange Q&A: p. 301) |
What converts mechanical energy to electrical energy? | Generator (Saia, Lange Q&A: p. 301) |
What is the primary function of the phototimer? | Terminate x-ray exposure once the IR is correctly exposed. (Saia, Lange Q&A: p. 282, 301) |
What is the voltage ripple of three phase, six pulse? | 13% (Saia, Lange Q&A: p. 283, 302) |
Which of the following will improve the spatial resolution of image-intensified images: 1. Very thing coating of cesium iodide 2. smaller diameter input screen 3. increased total brightness gain? | 1&2(Saia, Lange Q&A: p. 283, 202) |
What does a synchronous spinning top test look like? | Solid arc, with the angle representative of the exposure time (Saia, Lange Q&A: p. 284, 304) |
Star and Wye configurations are related to: | Three phase transformers (Saia, Lange Q&A: p. 284, 304) |
Automatic brightness control is used to adjust: | kV and mA (Saia, Lange Q&A: p. 284, 304) |
What is the device that directs the light emitted from the image intensifier to various viewing and imaging apparatus? | Beam Splitter (Saia, Lange Q&A: p. 285, 305) |
How often should kV settings be tested and how accurate should it be? | Annually, +/- 4 kV (Saia, Lange Q&A: p. 286, 307) |
A spinning top device can be used to evaluate: 1. timer accuracy 2. rectifier failure 3. effect of kV on contrast? | 1&2 (Saia, Lange Q&A: p. 287, 308 |
Total brightness gain of an image intensifier is the product of: 1. flux gain 2. minification gain 3. focusing gain? | 1 (flux gain) & 2 (minification gain) (Saia, Lange Q&A: p. 287, 208) |
Device used to change alternating currecnt to unidirectional current is? | Solid state diode (Saia, Lange Q&A: p. 287, 309) |
Deposited vaporized tungsten on inner surface of x-ray tube glass window does what to the tube/ beam? | Adds filtration to window (Saia, Lange Q&A: p. 288, 309) |
What device is used to control voltage by varying resistance? | Rheostat (Saia, Lange Q&A: p. 292, 314) |
What is rheostate? | Variable resistor that is used to change voltage or current values. It is found in the filament circuit. (Saia, Lange Q&A: p. 292, 314) |