SPC Mech Vent Alarms
SPC Mechanical Ventilation Alarm Settings
Question | Answer |
---|---|
Pressure Sensitivity Trigger | -1 to -2 |
Flow Triggering | -5 L/m / -2 to -3 L/m is common |
Low Pressure Alarm | 10cmH2O below Peak Pressure |
High Pressure Alarm | 10cmH2O above Peak Pressure |
Low Peep/CPAP Alarm | 2-3cmH2O below set level |
High Respiratory Rate Alarm | 10-15 above set number |
Low Respiratory Rate Alarm | 10 below set number |
Apnea Alarm | No more than 20 secs |
Spontaneous Rate Alarm | 10 above pt's spont brths |
Low Vt Alarm | 3 or 4 less than Ve |
High Ve Alarm | 20 above set Ve |