click below
click below
Normal Size Small Size show me how
Instrumentation
Instruments Chapter 3
Question | Answer |
---|---|
compound microscope | Light microscope |
The lens system consist of objectives and oculars | light microscope |
provide image maginification and image resolution | Objective lens |
Scanning lens, intermediate lens, high powered dry lens and immersion lens | types of objective lens |
will give an image surrounded by color fringes | Uncorrected lens also know as chromatic aberration |
are corrected for two colors red and blue and are found on most lab microscopes | achromatic objectives |
are corrected for three colors and for other lens aberrations | Apochromatic objectives |
are more expensive and not necessary for routine use. used for photomicrography | Apochromatic objectives |
a defect in lens system | aberration |
eyepiece | oculars |
have a x10 mag and x5 oculars are frequently used on student microscope x15 oculars preferred by microscopist | oculars |
is determined by multiplying the mag of ocular w/the objective | total mag obtained w/a microspe |
substage found below can be moved up/down | substage |
on modern microscopes consists of a condenser and iris diaphragm | substage |
function primarily to concentrate light on the tissue section | condenser |
should be centered accurately | condenser |
regulated by iris diaphragm and should be varied w/t different objectives | amount of illumination |
should be adjusted so that peripheral light rays are blocked and light passing through the tissue should be limited so that it fills the front lens of the objective | iris diaphragm |
keep covered when not in use, clean lens w/lens paper, remove immersion oil immediately after use | microscope maintenance |
use xylene on the ofectives as a last resort, dont dismantle objectives, | microscope maintenance |
when using immersion oil becareful, make sure that the high powered dry lens inst dragged through the oil | microscope maintenance |
reduce light t/a minimum or turn it off when not in use. remove slides from stage when not in use | microscope maintenance |
always focus upwards, never downwards, especially w/t higher power objectives | microscope maintenance |
do not touch lens surfaces w/fingers | microscope maintenance |
used in identification of crystals. Talc, silica or urates | polarizing microscope |
used in identification of amyloid stained w/congo red | polarizing microscope |
used to examine tissue for substances exhibiting the phenomena of double refraction, anisotropism and birefringence | polarizing microscope |
transmitting light unequally in different directions | birefringence |
having unlike properties in different directions | anisotropism |
is achieved by interposing a polarizing device between the light source and specimen and inserting a second polarizing filter between the specimen and the eye | polarized light |
vibrates in many planes | natural light |
vibrates only in one plane | light emerging from the polarizer |
can be converted to a polarizing microscope by placing one piece of polaring film on top of light source and another on top of the microscope slide. | light microscope |
through this microscope amyloid stained w/congo red, will show apple green birefringence | polarizing microscope |
used for examination of unstained specimens | phase-contrast microscope |
used to examine unstained living cells and allows transparent objects to be seen | phase-contrast microscope |
can be converted to a phase-contrast microscope by replacing the condenser and objective w/special phase equipment | a standard binocular microscope |
not used in routine histology | phase-contrast microscope |
directly transmitted light is excluded | dark field microscope |
only scattered or oblique light is utilize | dark field microscope |
objects appear much larger than they are, bcuz of their light scattering properties and frequently fine structures | dark field microscope |
used for the study of unstained microorganisms and is rarely used in routine histopathology | dark field microscope |
phenomeonon in whic light of one wavelenght is absorbed by as substance and instantly remitted as light of a longer wavelength | fluorescence microscope |
substance is bombarded w/short-wavelenght light in the ultraviolet, violet or blue range and visible light is emitted | fluorescence microscope |
two types transmission and scanning | electron microscope |
the specimen either transmits electrons producing electron lucent or clear areas in the image | transmission electron microscope |
the specimen deflects electrons (producing electron-dense, or dark area in the image | transmission electron microscope |
a two-dimensional black and white is seen on fluorescent screen | transmission electron microscope |
cells and ultrastructure of the cells can be appreciated w/ | transmission electron microscope |
useful in dx of kidney disease and in tumor identification | transmission electron microsocpe |
has replaced the electron microscope, in tumor diagnosis | immunoperoxidase |
a dramatic three-dimensional image | scanning electron microscope |
the electron beam sweeps the surface of the specime and releases secondary electrons | scanning electron microscope |
has a great depth of focus | scanning electron microscope |
very expensive, requires skilled operator | scanning electron microscope |
found in large instistutions | scanning electron microscope |
Rotary, sliding and freezing microtomes | most commonly used microtomes in histology |
used for sectioning glycol methacrylaten and paraffin embedded material | rotary microtome |
operates w/a screw feed | rotary microtome |
block moves up/dowm and either the knife or block advances a preset number of micrometers w/each revolution of the wheel | rotary microtome |
used for sectioning celloidin and large paraffin blocks | sliding microtome |
block held stationary and the knife is moved along a horizontal plane past the block face | sliding microtome |
not used in routine histology | sliding microtome |
has been replaced by the cryostat | clinical freezing microtome |
free-floating sections for some special stains are easier to obtain with this than with the cryostat | clinical freezing microtome |
a sharp knife must have and edge free of defects, is essential to obtain a good section | microtome knifes |
successful sectioning of poorly processed tissue is possible w/a | good knife |
nondiagnostic sections m/b obtained w/a | poor knife |
used for cutting plastics | glass knifes |
used to section glycol methacrylate embedded materials | ralph knifes |
used for sectioning, paraffin, celloidin, carbowax and frozen tissue | microtome knives |
made from high-grade tempered steel and are wedge-shaped, biconcave or planoconcave | microtome knives |
used for paraffin, carbowax, and frozen sections | wedge-shaped knife |
preferred for celloidin | planoconcave knife |
requires both block and knife to be kept wet | the wet celloidin method |
planoconcave knife can be used for this method | the wet celloidin method |
used for early hand sectioning | bioconcave knife |
has been replaced by the plane wedge knife | bioconcave knife |
rarely used | bioconcave knife |
wedge angle 15 to 18 | microtome knife |
27 to 32 | bevel angle |
condition of knife is critical, c/b determined w microscopic examination w/x100 mag | mictrome knife |
are stainless steel razor blades that are held secure in a holder on the microtome | disposable blades |
used to cut ultrathin sections for EM | diamond knifes |
should be discarded in punture-proof containers | disposables blades and glass knifes |
if used in cryostat, they need to be autoclaved before disposal | disposable blades |
knife tilt | clearance angle |
when cutting sections, angle should be 3 to 8 | clearance angle (knife tilt) |
should be determined with each knife | clearance angle (knife tilt) |
c/b corrected by adjusting the knife so that the correct clearance angle between knife and specimen is obtained | irregular, skipped or thick and thin sections |
horizontal edges of block are not parallel | ribbons are crooked |
mictoromy artifact | holes in sections |
should be performed less aggresively | block facing |
when holes appear in ribbons, an there is sufficient tissue in block, | ribbons should be cut until holes disappear |
dull knife | ribbon will not form |
depends on enough heat being generated by friction occurring as each section is cut to cause the sections to adhere to each other | ribbon formation |
caused by dull knife or too little knife tilt | section lifts from the knife as the block is raised |
warm room or paraffin to soft | section lifts from the knife as the block is raised |
occurs in very hard tissue like uterus | washboarding or undulations |
occurs when a block or knife has not been tightly clamped | washboarding or undulations |
too much knife tilt or worn microtome parts | washboarding or undulations |
overdehydration or lack of moisture | chatter, microscopic vibration appear in sections |
treating faced block for a few sections with wet cotton, or dip thumb in flotation bath and rugh the block | to remove chatter, or microscopic vibration |
dull knife, too much tilt, or cutting to rapidly | microscopic chatter |
too little knife tilt | sections are skipped or vary in thickness |
dull knife or gummed w/paraffin | sections are compressed, wrinkled or jammed |
too little knife tilt, too rapid cutting or too warm room | sections are compressed, wrinkled or jammed |
keep free of paraffin by wipping up w/a gause damped w/xylene (nerver wipe down | knife edge |
defect in knife | lenghtwise scratches or splits in ribbons |
static electricity, add moisture to air by breating on block and knife or boil water in open pan near microtome. | section flies or sticks to nearby objects |
refrigerated chamber containing a microtome (rotary) | cryostat |
cooled by mechanical regrigeration | cryostat |
sections are picked up directly onto slide | cryostat |
knife m/b sharp and edge free of defects | cryostat |
wedge-shaped knife is commonly used. | cryostat |
antiroll devices are designed for this knife | wedge-shaped knife |
operated at -20c | cryostats |
slow freezing allows water present in the tissue to form | ice crystals |
are seen as artifacts | ice crystal formation |
open and closed | two types of processors |
tissue is transported from one solution to the next | open processors |
tissue is stationary and fluids are pumped in and out of the chamber | closed processors |
are computerized w/a digital readout to indicate the exact instrument status | closed processors |
major advantage is that specimen cannot dry out w/i the chamber | closed processors |
disadvantage is that not all reagents c/b used on these processors | closed processors |
mercury, dichromate containing fixatives and chloroform, cant b/used on this processor | closed processors |
heat and vacumm | closed processors |
speed up processing | heat and vacumm |
use carefully on bx specimens | heat and vacumm |
m/b kept clean and a routine reagent rotation | processors |
fluid levels m/b higher than tissue | processors |
temp of paraffin m/b carefully adjusted to no more than 2C to 4C above melting point | processors |
temp m/b monitored and recorded daily | processors |
used to float out ribbons | flotation baths |
maintained at 5C to 10C below the melting point of paraffin used for embedding | flotation baths |
ribbons should be stretched gently, small wrinkles and folds should be teased out | flotation baths |
a mixture of egg white and glycerin w/elmers glue | albumin |
gellatin, agar and elmers glue | water bath additives |
poly-L lysine-coated slides | used for frozen sections, some to be stained in microwave and paraffin sections to be stained w/immunoenzyme technique |
postive charged attracts frozen and paraffin tissue sections elctrostatically | positive charged slides |
no adhesive needed | positive charged slides |
background staining is eliminated and section loss w/staining procedures is reduced | positive charged slides |
fairly expensive and not recommended for routine work | positive charged slides |
dissolve 1g of gelatin in 1L of distilled water w/heat. cool and add.1g of cromium potassium sulfate, store in fridge. | subbed slides |
after section is mounted on slide | it should be dried in the oven, hot air dryer or warming plate |
slides are dried to | remove water |
white spots that can be seen in tissue when slides are removed from xylene | incomplete deparaffination |
maintain a temp just above the melting point of paraffin | dryers and ovens |
can be used for overnight drying of immunoperoxidase slides but temp must be kept below 60C | ovens |
are maintained at 37C | incubators |
used for enzyme reactions and for some special stains | incubators |
used for storing may of the reagents used in histology, enzyme histochem reagents, ihc reagents, and buffer solutions | freezers and refrigeratos |
maintained at about 4C and should not be allowed to reach a temp above 10C | The refrigerator |
-20C | Freezer temp |
for IHC surface markers of lymphoid tissue or estrogen/progesterone receptor the freezer should be at what temp | -70C |
are used for performing special stains, for fixation, processing, drying of sides and other procedures | microwave ovens |
is dependent on exposure of time, solution volume and oven wattage | heat generated by microwave |
linear,revolving and robotic | automatic stainers found in histo |
transfer slides from one container to the next w/t same time allowed in each container | linear stainer |
also dries slides | linear stainer |
operate on the same prinicple as open tissue processor | revolving stainer |
are the most flexible stainer | robotic stainers |
allow computerized programming w/ability to return slides to same container used in a previous step | robotic stainers |
vary in slide capacity but slides may be continously loaded | linear stainer |
are batch stainers, one batch of slides m/b finisher before the next one is started | robotic and revolving stainers |
used to measure acidity. | pH meters |
scale goes from 0 to 14 | pH scale |
defined as netgative power to which the number 10 m/b raised to express the moles per liter concentration of solutions hydrogen ions | PH |
has melted paraffin w/warm storage for embedding molds, small warming and chilling plates, and larger chiller plate for rapid chilling of embedded tissue blocks | embedding center |
paraffin is kept 2C to 4C above melting point of parffing used | embedding center |