click below
click below
Normal Size Small Size show me how
Intro Rad Pro week 3
Intro to Rad Protection week 3
Question | Answer |
---|---|
Digital imaging techniques are always applied to | computed tomography (CT). ultrasound. magnetic resonance imaging. (All of the above.) |
The latent image is formed on the____ for digital radiography. | radiation detector |
Each matrix pixel represents_____ in digital fluorscopy. | atomic number and mass density |
In digital radiography, the brightness of the image is determined by | pixel values |
Digital imaging could only be developed after advances were made in | microprocessors. semiconductor memory. |
Computed radiography developed in the 1970s with the use of ____ as the image receptor. | phosphor |
Scanned projection radiography was developed as a complement to____. | computerized tomography. |
What is the most recent development in digital radiography image receptors? | direct capture solid-state devices |
Digital radiography first began to be used in clinical settings in the_____. | 1980s |
Spatial resolution is improved with increased____ in digital radiography. | matrix size |
The number of visible shades of gray in a digital image is determined by___. | bit depth |
The dynamic range determines the degree of___ in the image | contrast resolution |
_____ requires the largest matrix size. | Computed radiography |
The typical matrix size for digital radiography is____. | 1024 X 1024 |
Spatial resolution is determined by the ratio of____to____. | field of view, matrix size |
Scanned projection radiography differs from conventional radiography in the lack of | scattered x-rays. |
In scanned beam radiography the spatial resolution is determined by the | number of detectors. |
The primary limitation of fanned beam radiography is | long scanning time. |
Computed radiography utilizes____ for latent image formation. | photostimulable phosphor |
Conventional x-ray tubes and cassettes are used with____ radiography systems. | computed |
Detectors acquire info by | Scanning array detection |
The window level controls | image density |
The window width controls | image contrast |
Resolution is controlled by ____ size. | Pixel |
Detectors currently used in CR include all of the following except | capacitor plates |
Digital radiography replaces | Traditional film |
Silicon and selenium receptors are also known as | Flat panel detectors |
Which type of system uses a a two-part process to convert incoming x-ray photons to an electronic digital signal | indirect conversion |
During the two part conversion system a ___ converts light into an electronic digital signal | photodectector |
During the two part conversion system a _____ converts x-ray photons to light | Scintillator |
A photostimulable imaging plate includes a ___ layer | Phosphor, conductor, protective |
DR cassettes are also know as | Filmless cassettes |
The most common phosper use for computed radiography imaging plates is | barium fluorohalide bromides and iodides with europium activators |
The active layer of the CR imaging plate is made of | phototimulable fluorohalides |
The latent image is created by energy transfer during | photoeletric interactions |
The latent image will lose about 25% of its energy in | 8 hours |
The latent image is processed by a(n) | image reader device |
The CR imaging plate is scanned by a(n) | Helium-neon laser beam |
When the CR imaging plate is scanned the phosphors of the latent image release energy in the form of | Light |
During CR imaging plate processing, the laser frees trapped electrons allowing them to return to a lower energy state. This process is known as. | photostimulated luminescence |
Fluoro was developed so that radiologist could view ____ images | dynamic |
What is the milliamperage used during fluoro | 5mA |
The image intensifier improved fluoro by increasing image____. | brightness |
With image intensification the light level is raised to ____ vision. | photopic |
X-rays that exit the patient and enter the image intensifier first interact with the | input phosphor |
The output phosphor of the image intensifier is composed of___. | zinc cadmium sulfide |
The input phosphor converts___ to ___. | x-rays, light |
The ___ in the image intensifier emits electrons when it is stimulated by light photons | photocathode |
The number of light photons emitted withing the image intensifier is ___ to the amount of x-ray photons exiting the patient. | directly proportional |
Electrons hit the ____ after exiting the anode. | output phosphor |
The ___ is the product of the minification gain and the flux gain. | brightness gain |
The ratio of x-rays incident on the input phosphor to light photons exiting the output phosphor is called ___ gain. | flux |
The capability of an image intensifier to increase the illumination level of the image is called its. | brightness gain |
An image intensifier tube is identified by the diameter of its | input phosphor |
Brightness gain is typically in the range of___. | 5,000 - 30,000 |
Fluoro for an air contrast barium enema is generally done at ___ kV. | 80-90 |
Viewing the fluoro image in magnification mode increases | (all of the above) Contrast resolution, spatial resolution and patient dose |
Automatic brightness control(ABC) maintains the brightness of the image by varying | kV and mA |
The fluoro x-ray tube and image receptor are mounted on a | C arm |
The carriage commonly support the | (all of the above) image intensifier, power drive controls, and spot film selector |
During fluoro, the SOD cannot be less than ___ inches | 15 |
The photocathode absorbs ___ and emits ____. | light photons, electrons |
Electrostatic lenses are used to accelerate and focus | electrons |
The primary ___ occurs from the acceleration and focusing of the electron beam. | brightness gain |
The input screen is ____ in shape. | concave |
The shape of the input screen helps to control | distortion |
The photocathode is composed of ___ metals | photoemissive |
Photoemissive materials absorb ____ and emit ___. | light photons, electrons |
The greater the voltage supplied to the electrostatic lenses, the ___ the acceleration and the ___ the focal point to the input screen. | greater, closer |
the most common solution for quantum mottle is to | increase the fluoro tube mA. |
An electron that is described as being metastable: | Has higher energy than it should have |
Doping of a photostimulable phosphor with Europium results in | More stimulable light emission |
Photostimulable phosphor image receptors are effective because | Metastable states are produced |
Photostimulable phosphor image receptors | Can be fogged by background radiation |
To remove the image background radiation or a previous image, one should | Expose the image receptor to intense light |
The slow-scan portion of the computed radiography reader | Is under mechanical control |
Spatial resolution in computed radiography is principally determined by | laser beam diameter |
The source of the stimulating light is: | The laser |
How does the computed radiography reader maintain the laser beam as a circle | Beam-shaping optics |
The x-ray capture element of a computed radiography imaging plate is the | BaFlBr (w/europrium) |
An advantage of computed radiography over film/screen, regardless of the type of exam, is the relatively constant: | Image contrast |
Which of the following is an advantage of CR over film/screen. | Fewer repeats |
Digital radiography | Requires computer processing |
A scanned projection radiograph | Is virtually scatter-free |
The image receptor in CR is | BaFlCl |
Which of the following is used as a radiation detector in a DR | Selenium (silicon) |
The max frame acquisition rate in digital fluoro is about: | 30 frames per second |
During digital fluoro | The x-ray beam is pulsed |
How many video monitors are required for digital fluoro | Two |
Interrogation time: | Is the time needed to switch on the x-ray tube |
An advantage of digital imaging over conventional imaging is which of the following | postprocessing |
The mask image is usually the: | Image just preceding contrast |
A misregistration artifact | Occurs when a patient moves |
Remasking | Can correct for patient motion |
CR Computerized Radiography | Barium Fluorohalide with Europium (Cassette w/ IR) |
Indirect DR (Digital Radiography) | Digital Radiography where photons are converted to light then to electrons |
Indirect DR (Digital Radiography) | CsI or Gadolinium oxysulfide GdOS |
Indirect DR (Digital Radiography) | Thin film transitor or amorphous silicon convert electrons captured to light |
Direct DR | Involves no scintillating crystals |
Direct DR | Uses amporphous Selenium as capture and coupling element |
Minification times flux gain equals | Brightness gain |